Determination of reduced glutathione using an amperometric carbon paste electrode chemically modified with TTF–TCNQ
详细信息查看全文 | 推荐本文 |
摘要
The development of an amperometric sensor for the determination of reduced glutathione (GSH) is described. The sensor is based on tetrathiafulvalene–tetracyanoquinodimethane (TTF–TCNQ) incorporated into the graphite powder/Nujol oil matrix. The electrooxidation of GSH was monitored amperometrically at 200mV versus SCE (saturated calomel electrode). The amperometric response of the sensor was linearly proportional to the GSH concentration between 20 and 300μmoll−1, in 0.1moll−1 phosphate buffer (pH 8.0), containing 0.1moll−1 KCl and 0.5mmoll−1 Na2H2EDTA, as supporting electrolyte.