Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table
详细信息查看全文 | 推荐本文 |
摘要
Purpose: To scrutinize the positioning accuracy and reproducibility of a commercial hexapod robot treatment table (HRTT) in combination with a commercial cone-beam computed tomography system for image-guided radiotherapy (IGRT).

Methods and Materials: The mechanical stability of the X-ray volume imaging (XVI) system was tested in terms of reproducibility and with a focus on the moveable parts, i.e., the influence of kV panel and the source arm on the reproducibility and accuracy of both bone and gray value registration using a head-and-neck phantom. In consecutive measurements the accuracy of the HRTT for translational, rotational, and a combination of translational and rotational corrections was investigated. The operational range of the HRTT was also determined and analyzed.

Results: The system performance of the XVI system alone was very stable with mean translational and rotational errors of below 0.2 mm and below 0.2°, respectively. The mean positioning accuracy of the HRTT in combination with the XVI system summarized over all measurements was below 0.3 mm and below 0.3° for translational and rotational corrections, respectively. The gray value match was more accurate than the bone match.

Conclusion: The XVI image acquisition and registration procedure were highly reproducible. Both translational and rotational positioning errors can be corrected very precisely with the HRTT. The HRTT is therefore well suited to complement cone-beam computed tomography to take full advantage of position correction in six degrees of freedom for IGRT. The combination of XVI and the HRTT has the potential to improve the accuracy of high-precision treatments.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700