摘要
This paper reports the stable determination of l-methionine (L-Met) in phosphate buffer solution (pH = 4.0) using the electropolymerized film of 1,8,15,22-tetraaminophthalocyanato-copper(II) (p-4伪-CuIITAPc) on glassy carbon electrode (GCE). Bare GCE shows an ill-defined oxidation wave around 1.32 (卤 0.02) V for L-Met whereas p-4伪-CuIITAPc electrode shows a well-defined oxidation peak at 1.26 (卤 0.02) V. The modified electrode does not show any fouling effect towards the oxidation of L-Met. Further, the p-4伪-CuIITAPc modified electrode was successfully used for the selective determination of L-Met in the presence of ascorbic acid (AA) and uric acid (UA). Differential pulse voltammetry method was used to determine L-Met in the concentration range of 50-500 渭M with a detection limit of 2.7 脳 10鈭?#xA0;8 M (S/N = 3). The present modified electrode shows good recovery results for spiked L-Met in human blood serum sample.