Patient motion compensation during transthoracic 3-D echocardiography
详细信息查看全文 | 推荐本文 |
摘要
Bulk patient motion during transthoracic 3-D echocardiography (3DE) produces image plane misregistration and errors in left ventricular (LV) volume and ejection fraction (EF). To correct for patient motion, we used a magnetic locating system to track both the ultrasound transducer and the chest wall of the patient, so images could be registered in a patient-centered coordinate system (“correction”). Fourteen subjects each underwent 3DE, with deliberate patient motion, to measure LV volume and EF. Results were compared to magnetic resonance imaging (MRI). Without correction, 3DE differed significantly from MRI (EF: r = 0.78, SEE = 5.8%). Application of correction increased 3DE accuracy, despite patient motion (EF: r = 0.91, SEE = 3.7%), to a level comparable to that of 3DE in the absence of motion (EF: r = 0.93, SEE = 3.5%). Patient motion during 3DE examination can be corrected using a magnetic spatial location system. (E-mail: psd@medicine.wisc.edu)