Validaci贸n del dispositivo port谩til COPD-6 para la detecci贸n de patolog铆as obstructivas de la v铆a a茅rea
详细信息查看全文 | 推荐本文 |
摘要

Objectives

To evaluate the agreement and the association with FEV1, FEV6 and FEV1/FEV6 measured with the Vitalograph-COPD-6 portable device and the FEV1, FVC and FEV1/FVC by conventional spirometry, and to analyse the validity of this device to detect obstruction.

Methodology

A cross-sectional, descriptive, prospective study, that included 180 subjects. A conventional spirometry and one with the Vitalograph-COPD-6 were sequentially performed on them. The agreement was analysed [kappa index and interclass correlation coefficient (ICC)], as well as the association [Pearson correlation coefficient (r)], area under the ROC curve (AUC) of the FEV1/FEV6 in detecting obstruction, and the sensitivity, specificity, predictive values (PPV and NPV), and probability ratios (PR+ and PR-) of the different FEV1/FEV6 cut-off points in the detection of obstruction.

Results

The prevalence of obstruction was 47%. The kappa index was 0.59 when an FEV1/FEV6 < cut-off point of <0.7 was used. The ICC and the r between the FEV1 measured by the two instruments, FEV6 and FEV1/FEV6 measured by the Vitalograph-COPD-6 and the FVC and FEV1/FVC determined by the spirometer were all greater than 0.92. The ROC AUC was 0.97. To detect obstruction, if the cut-off point of FEV1/FEV6 (for COPD-6) was <0.70, the sensitivity, specificity, PPV, NPV, CR+ and CR鈭?were, 58%, 100%, 100%, 73%, 鈭?and 0.42, respectively. For a cut-off point of <0.8, they were 96%, 76%, 78%, 96%, 3.8 and 0.05, respectively.

Conclusions

The portable Vitalograph-COPD6 device is precise for the detection of airway obstruction. The best sensitivity/specificity of FEV1/FEV6 was obtained with cut-off points greater than 0.7.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700