Absorption tomography of laser induced plasmas
详细信息查看全文 | 推荐本文 |
摘要
An emission tomography of laser-induced plasmas employed in the laser induced breakdown spectroscopy (LIBS) requires signal integration times in a microsecond range during which the LIBS plasma cannot be considered stationary. Consequently, the use of the data for reconstructing the plasma properties under the assumption that the latter does not change significantly during the integration time leads to inaccurate results. To reduce the integration time, it is proposed to measure a plasma absorption in parallel rays using a scanning rectangular aperture whose dimension along the scanning direction is about a characteristic size of plasma plumes () and the other dimension is of the order of a uniformity length of plasma parameters (). The aperture is moved step by step along the scanning direction and the total energy of photons coming through the aperture is measured during time T at each position of the aperture. Owing to the large size of the aperture, the integration time T is reduced by a factor . A numerical data processing is proposed to restore the spatial resolution of the plasma absorption along the scanning direction. It is determined by the scanning step . Another advantage of the proposed procedure is that inexpensive linear CCD or non-discrete (PMT, photodiode) detectors can be used instead of costly 2-dimensional detectors.