Retrospective Dosimetric Comparison of Low-Dose-Rate and Pulsed-Dose-Rate Intracavitary Brachytherapy Using a Tandem and Mini-Ovoids
详细信息查看全文 | 推荐本文 |
摘要
The purpose of this study was to compare the dose distribution of Iridium-192 (192Ir) pulsed-dose-rate (PDR) brachytherapy to that of Cesium-137 (137Cs) low-dose-rate (LDR) brachytherapy around mini-ovoids and an intrauterine tandem. Ten patient treatment plans were selected from our clinical database, all of which used mini-ovoids and an intrauterine tandem. A commercial treatment planning system using AAPM TG43 formalism was used to calculate the dose in water for both the 137Cs and 192Ir sources. For equivalent system loadings, we compared the dose distributions in relevant clinical planes, points A and B, and to the ICRU bladder and rectal reference points. The mean PDR doses to points A and B were 3%± 1%and 6%± 1%higher than the LDR doses, respectively. For the rectum point, the PDR dose was 4%± 3%lower than the LDR dose, mainly because of the 192Ir PDR source anisotropy. For the bladder point, the PDR dose was 1%± 4%higher than the LDR dose. We conclude that the PDR and LDR dose distributions are equivalent for intracavitary brachytherapy with a tandem and mini-ovoids. These findings will aid in the transfer from the current practice of LDR intracavitary brachytherapy to PDR for the treatment of gynecologic cancers.