ALP catalytic activity was measured in 50 serum samples and 16 commercial materials, including control materials from EQAS programs, using primary reference measurement procedure and two routine measurement procedures with AMP and DEA as buffers. Calibration materials with a value assigned by reference procedure which were proved to be commutable were used to recalculate the serum values obtained by routine procedures.
All commercial materials showed a similar behaviour to the patient specimens when AMP vs IFCC procedures were compared. For DEA vs IFCC comparison only one calibration material and two quality control materials were commutable. Recalculation of serum results with a commutable common calibrator improves the agreement between methods changing the ratio AMP vs IFCC from 1.44 to 1.04 and DEA vs IFCC from 3.02 to 1.05.
The use of a common commutable calibration material allows harmonizing ALP measurements and made traceable patient results to reference procedure.