k0-NAA quality assessment in an Algerian laboratory by analysis of SMELS and four IAEA reference materials using Es-Salam research reactor
详细信息查看全文 | 推荐本文 |
摘要
Different types of synthetic multi-element standard material (SMELS) and four IAEA reference materials, 140, Sl-1, Soil-7 and Lichen-336 were analyzed for validation and QC/QA of the k0-standardised Neutron Activation Analysis (k0-NAA). The samples of SMELS and RMs were irradiated at Es-Salam research reactor and measured on an absolutely calibrated HPGe detector with 35%relative efficiency connected to a Canberra Genie 2k inspector. Concentrations of 33 elements such as As, Au, Ba, Br, Ca, Ce, Co, Cr, Cs, Eu, Fe, Hf, In, K, La, Mn, Mo, Na, Nd, Rb, Sb, Sc, Se, Sm, Sr, Ta, Tb, Th, Tm, U, Yb, Zn, and Zr were determined in SMELS and RMs. The analytical results agreed well with the assigned values of SMELS and certified values of RMs. In the case of RMs, concentrations of a few elements, whose certified values are not available, could be determined. The comparison between experimental values and assigned/certified data for SMELS and RMs was made by means of the results from Relative Bias, Z-score and U-score. The relatives bias of the elements determined in SMELS with respect to the assigned values were all within卤4.6%. For RMs with respect to certified values were within卤10%except for few elements for which RB varied from 鈭?8.6%to +12.8%. The Z-score values at 95%confidence level for most of the elements in both the materials were within 卤1. The U-scores for most of the elements were lower than 1.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700