The aCRT algorithm is currently undergoing evaluation in a prospective, randomized, double-blinded, worldwide clinical trial. The trial enrolled 522 patients, who satisfied standard clinical indications for a CRT device. Within 2 weeks after the implant, the patients were randomized to aCRT versus echo-optimized biventricular pacing (Echo) settings in 2:1 ratio and followed up at 1-, 3-, 6-, and 12-month postrandomization. The noninferiority primary trial objectives at 6-month postrandomization are to demonstrate that (a) the percentage of aCRT patients who improved in their clinical composite score is at least as high as the percentage of Echo patients; (b) cardiac performance as assessed by echocardiography is similar when using aCRT settings versus echo-optimized settings; and (c) aCRT does not result in inappropriate device settings. First and last patient enrollments occurred in November 2009 and December 2010, respectively.
The safety and efficacy of the aCRT algorithm will be evaluated in this ongoing clinical trial.