Assessment of reduced glutathione: Comparison of an optimized fluorometric assay with enzymatic recycling method
详细信息查看全文 | 推荐本文 |
摘要
Glutathione is an important tripeptide involved in a variety of cellular processes. Thus, precise knowledge of its levels is essential. Glutathione exists in two free forms鈥攔educed and oxidized鈥攁nd a number of methods exist to measure its levels. The aim of our work was to optimize a spectrofluorometric assay for reduced glutathione based on the reaction between glutathione and o-phthalaldehyde. We found that a change of excitation wavelength to 340 nm and modification of pH to 6.0 enhance sensitivity and specificity of the method (intraassay coefficient of variation CV <3%, interassay CV = 5.1%, recovery = 98-102%, linearity = 0-1000 渭M GSH, calibration R2 = 1.00). We also anticipated possible effect of various amino acids on the fluorescence signal, but no interference was found. We compared the optimized fluorometric method with a popular enzymatic recycling glutathione assay and found very strong correlation of results (= 0.99, = 45). We introduce here an optimized fluorometric method possessing sufficient sensitivity and specificity that is comparable to the enzymatic glutathione assay. Because the fluorometric assay procedure is faster and lower in cost, it could be ideal for routine analysis of reduced glutathione levels in a large number of samples.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700