New applications of the VDmax approach to substantiation of preselected sterilization doses
详细信息查看全文 | 推荐本文 |
摘要
The VDmax approach for substantiation of 25 kGy has been in use for more than 10 years. VDmax methods are included in ISO 11137-2:2006 and . ISO Technical Specification, 13004, is under development that will include sterilization doses from 15 to 35 kGy.

For substantiation of a sterilization dose for very low bioburden products, less than or equal to 0.3, values of VDmax have now been derived and tabulated for a sterilization dose of 12.5 kGy.

Products have been encountered that have both low bioburden and a relatively low maximum dose. In several situations, existing tabulated VDmax values could not be effectively used; in one such situation, the average bioburden was too high to substantiate a 15 kGy sterilization dose and the use of a 17.5 kGy sterilization dose was not practicable due to the likelihood of exceeding the product's maximum acceptable dose. For this product, values of VDmax were derived and tabulated for substantiation of a 16.1 kGy sterilization dose.

Values of VDmax have been derived and tabulated for the substantiation of sterilization doses linked to a sterility assurance level (SAL) of 10鈭?. To offer a potential alternative to aseptic processing, the notion of using an 鈥渁septic processing equivalent dose鈥? 10鈭? SAL, has been investigated along with the use of alternate model populations for calculation of VDmax values.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700