Trace analysis of residual methyl methanesulfonate, ethyl methanesulfonate and isopropyl methanesulfonate in pharmaceuticals by capillary gas chromatography with flame ionization detection
详细信息查看全文 | 推荐本文 |
摘要
A capillary gas chromatographic method using flame ionization detection was developed and validated for the trace analysis (ppm level) of methyl methanesulfonate, ethyl methanesulfonate, and isopropyl methanesulfonate in pharmaceutical drug substance. The method utilizes a megabore capillary column with bonded and crosslinked polyethylene glycol stationary phase. A dissolve-and-injection approach was adopted for sample introduction in a splitless mode. The investigated sample solvents include acetonitrile, ethyl acetate, methylene chloride, 1,2-dichloromethane, and toluene. Aqueous mixtures of acetonitrile and water can also be used as sample solvent. A limit of detection of about 1μg/g (1ppm) and limit of quantitation of 5μg/g (5ppm) were achieved for the mesylate esters in drug substance samples. The method optimization and validation are also discussed in this paper.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号