Volatile and semivolatile organo-halogen trace analysis in surface water by direct aqueous injection GC-ECD
详细信息查看全文 | 推荐本文 |
摘要
A procedure for simultaneous determination of volatile and semivolatile chloroorganic compounds (from trichloromethane to di- and trichlorobenzenes to hexachlorobutadiene) in aqueous samples was developed. The procedure is based on a DAI-GC-ECD technique widely used for trihalomethanes determination in drinking water. The DAI technique application could have been extended to semivolatile chloroorganic compounds by using a Rtx-624 (60 m × 0.32 mm I.D., df 1.8 μm (6%cyanopropyl, 94%dimethyl polysiloxane). Precision (R.S.D.) and detection limits of the DAI-GC-ECD procedure with respect to 12 volatile and semivolatile chloroorganics were determined. Relative standard deviations (R.S.D.) were lower than 22%and the detection limits (concentration equivalent to a peak height three times the baseline noise) range from 0.02 μg L−1 for trichloroethene and tetrachloroethene to 0.25 μg L−1 for hexachlorobutadiene to 1.6 μg L−1 for 1,2-dichlorobenzene. The method was successfully used to determine chloroorganic pollutants in river water.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083