Compartmental Bone Morphometry in the Mouse Femur: Reproducibility and Resolution Dependence of Microtomographic Measurements
详细信息
下载全文
推荐本文 |
摘要
Microcomputed tomography (μCT) is widely used for nondestructive bone phenotyping in small animals, especially in the mouse. Here, we investigated the reproducibility and resolution dependence of μCT analysis of microstructural parameters in three different compartments in the mouse femur. Reproducibility was assessed with respect to precision error (PE%CV) and intraclass correlation coefficient (ICC). We examined 14 left femurs isolated postmortem from two strains of mice (seven per group). Measurements and analyses were repeated five times on different days. In a second step, analysis was repeated again five times for a single measurement. Resolution dependence was assessed by high-resolution measurements (10 μm) in one strain and subsequent image degrading. Reproducibility was better in full bone compartment and in cortical bone compartment in the diaphysis (PE%CV = 0.06–2.16%) than in trabecular compartment in the distal metaphysis (PE%CV = 0.59–5.24%). Nevertheless, ICC (0.92–1.00) showed a very high reliability of the assessed parameters in all regions, indicating very small variances within repeated measurements compared to the population variances. Morphometric indices computed from lower- and higher-resolution images displayed in general only weak dependence and were highly correlated with each other (R2 = 0.91–0.99). The results show that parameters in the full and cortical compartments were very reproducible, whereas precision in the trabecular compartment was somewhat lower. Nevertheless, all compartmental analysis methods were very robust, as shown by the high ICC values, demonstrating high suitability for application in inbred strains, where highest precision is needed due to small population variances.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700