摘要
针对RS6000流变仪在压裂液耐温耐剪切性能测试中遇到的温度控制不精确、剪切速率波动及粘度测量有偏差等问题,采用实验数据分析、测量单元测试原理分析及影响因素分析对RS6000流变仪测量过程中出现的上述问题进行了校准,实验结果显示温度误差在0~2.5℃之间,测量间隙Gap值在1~2.9 mm之间,受转子与轴承的磨阻影响的粘度值可采用"stationary flow curve (CS/CR-rotation step)"方法拟合参数a,b,c进行校正。校准后RS6000流变仪测量的数据稳定性得到明显改善。
The problems of temperature control imprecision, shear rate fluctuation and viscosity measurement deviation in the test of temperature and shear resistance are summarized.The above problems occurring in RS6000 rheometer measurement are calibrated by the experimental data analysis, measurement unit test principle analysis and influencing factor analysis. The experiment results show that the temperature error is between 0~2.5 ℃, the measuring Gap is between 1~2.9 mm, and the viscosity values influencingby rotor and bearing friction can be used "stationary flow curve(CS/CR-rotation step)" method fitting parameters a, b, c. The data stability of RS6000 rheometer is significantly improved through the calibration, which provides some enlightenment and reference for the calibration of the rheometer.
引文
[1] 李宗田,李凤霞,黄志文.水力压裂在油气田勘探开发中的关键作用的[J].油气地质与采收率,2010,17(5):76-79.
[2] 李文博.延长气田压裂液体系室内评价研究[J].石油化工应用,2014,33(12):118-120.
[3] 姜阿娜,仲岩磊,陈凯,等.羟丙基瓜胶压裂液流变行为影响因素研究[J].中国石油大学胜利学院学报,2011,25(1):15-17.
[4] 姜阿娜.醇基压裂液在深层低渗油藏压裂改造中的应用[J].精细石油化工进展,2012,13(7):5-7.
[5] 张翠林.MCR301流变仪使用经验[J].新疆石油科技,2002,22(1):44-46.
[6] 夏宏,张海龙,王珂昕,等.温度对无伤害压裂液性能影响初探[J].石油天然气学报(江汉石油学院学报),2006,28(4):357-358.
[7] 包玲,刘家林,曲波,等.流变仪与黏度计测定聚合物黏度的影响因素研究[J].精细石油化工,2012,13(10):10-12.
[8] 祁海洋.高压旋转粘度计的研制及合成油压粘特性研究[D].哈尔滨:哈尔滨工业大学,2011.