摘要
目的对BC-5310血细胞分析仪性能进行评价。方法使用新鲜EDTA-K2抗凝全血及厂家所供配套校准品、质控品,根据NCCLS有关规定的指标对该仪器进行测试评价。结果各项指标的批内精密度、携带污染率、可比性、总重复性等符合标准要求。结论 BC-5310血细胞分析仪精密度高、携带污染率低、重复性好,综合性能优良,适合医院常规、急诊或健康体验使用。
Objective To evaluate performance of BC-5310 blood cell analyzer. Methods Carry on test and evaluation for instrument with fresh EDTA-K2 anticoagulant whole blood and matching calibration and quality control products provided by manufacturer according to relevant indicators of NCCLS. Results Batch accuracy, contamination rate, comparability and total repeatability of indicators meet requirements of standard. Conclusion BC-5310 blood cell analyzer has high precision, low contamination rate, good repeatability and excellent comprehensive performance,which is suitable for routine, emergency or health checkup of hospitals.