高压蒸汽灭菌包内化学指示剂与PCD批量监测效果研究
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:Batch monitoring efficacy of chemical indicators and PCD in high-pressure steam sterilization
  • 作者:吴金京 ; 巴宁 ; 王红光 ; 邱翠华 ; 李志明 ; 王倩 ; 施慧 ; 周丽君
  • 英文作者:WU Jin-jing;BA Ning;WANG Hong-guang;QIU Cui-hua;LI Zhi-ming;WANG Qian;SHI Hui;ZHOU Li-jun;Nanjing General Hospital of Nanjing Command;
  • 关键词:压力蒸汽灭菌 ; 化学指示剂 ; 过程挑战检测装置 ; 监测效果
  • 英文关键词:Pressure-steam sterilization;;Chemical indicator;;PCD;;Monitoring efficacy
  • 中文刊名:ZHYY
  • 英文刊名:Chinese Journal of Nosocomiology
  • 机构:南京军区南京总医院消毒供应科;广州军区武汉总医院保健科;南京军区南京总医院信息科;
  • 出版日期:2015-12-02 11:02
  • 出版单位:中华医院感染学杂志
  • 年:2015
  • 期:v.25
  • 基金:解放军重大基金资助项目ANJ13JJ001;; 南京军区医学科技创新基金资助项目(14ms101)
  • 语种:中文;
  • 页:ZHYY201524086
  • 页数:3
  • CN:24
  • ISSN:11-3456/R
  • 分类号:262-264
摘要
目的调查两种化学指示剂在高压蒸汽灭菌中的监测效果,了解两种化学指示剂灵敏度,确保消毒灭菌效果,保障医院平战时消毒灭菌物品供应。方法于2014年9月选择了器械包、敷料包、应急采血物资包、铺巾包各50个,每个包在同一部位放置两种3M化学指示剂和1种3M生物指示剂,分5锅次灭菌完成,对比灭菌合格率,数据采用SPSS 17.0软件进行统计分析。结果 1250多参数包内化学指示卡检测器械包合格率为78.0%、敷料包为96.0%、应急采血物资包为94.0%、铺巾包为98.0%;1243A综合参数包内化学指示卡监测灭菌合格率100.0%;1292嗜热脂肪杆菌指示剂监测灭菌合格率100.0%;每锅次41360化学过程挑战检测装置(PCD)和1292生物测试包监测合格率100.0%。结论 PCD批量监测可以保证锅内所有灭菌物品安全合格;在紧急情况下,可以用含有第五类化学指示剂的PCD进行监测,提前放行植入型医疗器械。
        OBJECTIVE To observe the monitoring efficacy of chemical indicators and PCD in the high-pressure steam sterilization and understand the sensitivity of the two kinds of indicators so as to ensure the disinfection and sterilization efficacy and guarantee the supply of disinfection and sterilization materials in hospitals in peace time and war time.METHODS The monitoring efficacy was assessed in the sterilization of the four kinds of medical bags,including instrument sets,dressing bags,blood-collecting bags and gauze bags,with 50 bags of each kind.Two kinds of 3Mchemical indicators and one kind of 3M biological indicator were put in the same place in each bag.The bags were sterilized for sterilizing pots.The sterilization qualification rates were compared.The software SPSS 17.0 was used for statistical analysis.RESULTS The 1250 pack interior chemical indicative card showed that the sterilization qualification rate was 78.0% for instrument sets,96.0% for dressing bags,and98.0%for gauze bags.The 1243 Acomprehensive pack interior chemical indicative card showed that the sterilization qualification rates were 100.0%.The sterilization qualification rate was 100.0% for the monitoring of 1292 Bacillus stearothermophilus indicator.Chemical 41360 PCD and 1292 biological indicator showed that the sterilization quality rates were 100.0%.CONCLUSIONBatch monitoring of PCD can guarantee the safety and qualification of all sterilized materials in the pot.In emergency,PCD containing the five type chemical indicators can be used for monitoring and placed in implantable instruments in advance.
引文
[1]周婉蛾,卢玉珍,李彦兰,等.化学指示剂在不同物品灭菌中监测效果[J].当代医学2013,19(16):67-68.
    [2]黄靖雄,何珉,王之栋.压力蒸汽灭菌监测进展[J].中国护理管理,2012,12(7):12-14.
    [3]张正焘,吴伟.压力蒸汽灭菌过程挑战装置进展[J].中国护理管理,2011,11(5):71-72.
    [4]陈春莲,李玉平,赖秀梅.比较在不同的灭菌时间点生物指示剂与化学指示卡对压力蒸汽灭菌监测的效果[J].国际护理学杂志,2011,30(5):729-730.
    [5]姜华,屈云娜,艾芬,等.压力蒸汽灭菌化学监测测试包的使用观察与分析[J].中华医院感染学杂志,2008,18(12):1726-1727.
    [6]van Doornmalen JP,Dankert J.A validation survey of 197hospital steam sterilizers in The Netherlands in 2001and 2002[J].J Hosp Infect,2005,59(2):126-130.
    [7]刘龙秀,曾金凤.灭菌过程生物与化学监测与应用[J].中国民族民间医药,2012,21(10):34.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700