摘要
针对开放光路TDLAS(可调谐半导体激光吸收光谱技术)气体检测系统中,由于角反射器在生产及使用过程中表面光洁度下降及大气传输扰动造成回波信号入射方向与接收透镜光轴非严格平行,从而导致系统探测灵敏度下降的问题,分析回波信号与菲涅尔透镜光轴夹角不同时的汇聚点偏移量,设计了由圆锥体反射器和抛物面反射器组成的二次光学元件,并运用TracePro光学软件对其参数进行优化。在回波光线-6°~6°入射角偏差范围条件下对优化设计的光学接收组件进行聚光特性模拟。模拟结果表明:优化设计的光学接收组件接收角为4.9°,光学效率为78.24%,与单独菲涅尔透镜及菲涅尔透镜-复合抛物面聚光器(CPC)光学接收组件相比聚光性能明显提高,适用于实际检测过程。
In the open-optic path tunable diode laser absorption spectroscopy(TDLAS)detection system for gas,due to the surface finish degeneration of corner reflector during production and using and the atmospheric transmission disturbance,the incident direction of echo signal and the optical axis of receiving lens are not strictly parallel,resulting in the decrease of the system detection sensitivity.Aiming at this problem,firstly the offset of convergence point with the incident direction of echo signal different from the the optical axis of the Fresnel lens was analyzed,then a secondary optical element consisting of a conical reflector and a parabolic reflector was designed and its parameters were optimized by TracePro software.Moreover,the condensing performance simulation under the deviation range of echo incident angle in -6° ~ 6° was carried out.The simulation results show that the receiving angle of the optimized optical receiving module is 4.9° and the optical efficiency is 78.24% ,the condensing performance is obviously improved comparing to the independent Fresnel lens and Fresnel lens-compound parabolic concentrator(CPC),the optimized component is applicable to the actual detection process.
引文
[1]Karla M C,Armando S J,Nayeli E O,et al.Methanogenic activity tests by infrared tunable diode laser absorption spectroscopy[J].Journal of Microbiological Methods,2012,91(1):89-92.
[2]Bushmeleva K I,Plyusnin I I,Bushmelev P E,et al.Modeling the optimal parameters for a remote sensing device[J].Measurement Techniques,2011,54(3):294-299.
[3]Roller C,Fried A,Walega J.Advances in hardware,system diagnostics software,and acquisition procedures for high performance airborne tunable diode laser measurements of formaldehyde[J].Applied Physics B,2006,82(2):247-264.
[4]Van Well B,Murray S,Hodgkinson J,et al.An openpath,hand-held laser system for the detection of methane gas[J].Journal of Optics A:Pure and Applied Optics,2005,7(6):420-424.
[5]Iseki T.A portable remote methane detector using an InGaAs PDFB laser[J].Environmental Geology,2004,46(8):1064-1069.
[6]Yao Hua.Research on remote sensing of methane based on tunable diode laser absorption spectroscopy technique[D].Zhejiang:Zhejiang University,2011:16-55.姚华.采用可调谐激光吸收光谱技术遥测甲烷气体浓度的研究[D].浙江:浙江大学,2011:16-55.
[7]Taslakov M,Simeonov V,Froidevaux M,et al.Openpath ozone detection by quantum-cascade laser[J].Applied Physics B:Lasers and Optics,2006,82(3):501-506.
[8]Wang Liming,Zhang Yujun,Li Hongbin,et al.Effect of turbulence on laser absorption signal and its improvement method[J].Laser Technology,2012,36(5):670-673.王立明,张玉均,李宏斌.湍流对激光吸收光谱信号的影响及改善方法研究[J].激光技术,2012,36(5):670-673.
[9]Yang Guanghui,Wei Ming,Chen Bingchen,et al.Design and research of equal-thickness slab Fresnel lens[J].Journal of Applied Optics,2013,34(6):898-902.杨光辉,卫明,陈丙振,等.等厚型平板菲涅尔透镜设计与研究[J].应用光学,2013,34(6):898-902.
[10]An Yupeng,Wang Yiding,Li li,et al.Design of the optical structure of the mid-infrared gas sensing system by using the TracePro[J].Laser&Infrared,2009,39(11):1198-1202.安宇鹏,王一丁,李黎,等.采用TracePro进行中红外气体检测系统的光路结构设计[J].激光与红外,2009,39(11):1198-1202.
[11]Victoria M,Dominguea C,Anton I,et al.Comparative analysis of different secondary optical elements for aspheric primary lenses[J].Optics Express,2009,17(8):6487-6492.
[12]Stutz J,Platt U.Numerical analysis and estimation of the statistical error of differential optical absorption spectroscopy measurements with least-squares methods[J].Applied Optics,1996,35(30):6041-6053.