摘要
目的建立石墨炉原子吸收分光光度法测定阿胶中铅含量的不确定度分析方法,找出影响不确定度的因素。方法采用石墨炉原子吸收分光光度法测定阿胶中铅含量,并根据《测量不确定度评定与表示》(JJF1059.1-2012)有关规定评定其不确定度。结果阿胶中铅的含量为0.4660mg·kg~(-1),扩展不确定度为0.05686mg·kg~(-1)(k=2,置信概率95%)。结论不确定度主要由标准曲线拟合所引入。
OBJECTIVE To establish the uncertainty in the determination of plumbum in Asini corii colla by graphite furnace atomic absorption spectrometry,and to identify the factors that affected its uncertainty.METHODS Graphite furnace atomic absorption spectrometry was used to determine the content of plumbum in Asini corii colla and the uncertainty of the determination was estimated based on "Evaluation and Expression of Uncertainty in Measurement(JJF1059.1-2012)".RESULTS The plumbum in Asini corii colla was 0.4660 mg·kg~(-1),and expanded uncertainty was 0.05686 mg·kg~(-1)(k=2,confidence probability of 95%).CONCLUSION The uncertainty is mainly caused by standard curve fitting.
引文
[1]刘元涛,张惠惠,王升光,等.阿胶仿生酶解前后提高免疫力作用对比研究[J].时珍国医国药,2016,27(9):2158-2160.
[2]栾爽,赵迎春,韩春晖,等.77种中药材中铅、镉残留量的测定与分析[J].中国药房,2015,26(12):1678-1681.
[3]王蓓蓓.骨刺片中重金属及有害元素的控制[J].海峡药学,2016,28(3):71-73.
[4]杨江,董小萍.中药中重金属元素检测方法的研究进展[J].中药与临床,2014,5(4):46-49.
[5]邹义栩.ICP-MS法测定复方地龙胶囊中重金属元素的含量[J].海峡药学,2016,28(11):66-67.
[6]国家药典委员会编.中国药典[S].一部.北京:中国医药科技出版社,2015.
[7]国家质量技术监督局.JJF1059-2012测量不确定度评定与表示[S].北京:中国计量出版社,2012.
[8]国家质量监督检验检疫总局.JJG196-2006常用玻璃量器检定规程[S].2006,6-8.