自动读取CR-39固体核径迹探测器径迹数的简单方法
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:The Simple Method to Read the Number of Tracks from CR-39 Solid State Nuclear Track Detector Automatically
  • 作者:谭延亮 ; Shinji ; Tokonami ; 覃金宁 ; 管少斌 ; 何春明 ; 周宗杰 ; 刘旭 ; 李峰林 ; Paitoon ; Wanabongse ; Masahiro ; Hosoda ; 袁红志 ; Hiromi ; Kudo-Yokota ; Chanis ; Pornnumpa
  • 英文作者:TAN Yan-liang;Shinji Tokonami;QIN Jin-ning;GUAN Shao-bin;HE Chun-ming;ZHOU Zhong-jie;LIU Xu;LI Feng-lin;Paitoon Wanabongse;Masahiro Hosoda;YUAN Hong-zhi;Hiromi Kudo-Yokota;Chanis Pornnumpa;College of Physics and Electronics,Hengyang Normal University;Department of Radiation Physics,Institute of Radiation Emergency Medicine,Hirosaki University;Gelogical Bureau of Hunan Province Nuclear Industry Brigade 306;Airbome Survey and Remote Sensing Center of Nuclear Industry;
  • 关键词:CR-39 ; 径迹 ; 自动读取
  • 英文关键词:CR-39;;track;;automatic read
  • 中文刊名:HYSF
  • 英文刊名:Journal of Hengyang Normal University
  • 机构:衡阳师范学院物理与电子工程学院;弘前大学应急医学研究所;湖南省核工业地质局三〇六大队;核工业航测遥感中心;
  • 出版日期:2018-06-15
  • 出版单位:衡阳师范学院学报
  • 年:2018
  • 期:v.39;No.223
  • 基金:国家自然科学基金资助项目(11375058);; 衡阳市科技局产学研项目(2017KJ159)
  • 语种:中文;
  • 页:HYSF201803012
  • 页数:3
  • CN:03
  • ISSN:43-1453/Z
  • 分类号:54-56
摘要
人工使用显微镜对蚀刻后的CR-39观测读数很容易产生视觉疲劳,自动分析读数的机器价格昂贵。该文介绍了日本弘前大学应急医学研究所采用的低成本解决方案:使用底片扫描仪或数码相机通过显微镜拍照的方式将蚀刻后的CR-39探测器表面图像数字化;对数字化后的图像用免费的ImageJ图像处理软件进行分析就可以完成对CR-39的自动读片。自动读片与人工读片的差异在可接受的范围之内。
        Manual observation and reading the number of tracks of etched CR-39 through microscope results in vision fatigue easily,automated analysis and reading machines are expensive.In this paper,we introduce a low-cost solution from the Institute of Emergency Medicine,Hirosaki University,Japan.Digital images of the etched CR-39 detectors surface can be obtained by a film scanner or a digital camera with a microscope.Automatic readings the tracks of CR-39 can be achieved through the analysis of digital images by free image processing software,ImageJ.The difference between the automatic and manual reading is in the acceptable range.
引文
[1]CARTWRIGHT B G,SHIRK E K,PRICE P B.A nuclear-track-recording polymer of unique sensitivity and resolution[J].Nuclear Instrument and Methods,1978,153(2-3):457-460.
    [2]翟鹏济,唐孝威,王龙,等.CR-39核径迹探测器及其在核科学等研究领域中的应用[J].物理,2000,29(7):397-401.
    [3]胥建国,李航,邹宇斌,等.用组合叠层CR-39固体径迹探测器测量加速器厚铍靶9Be(d,n)反应中子能谱[J].核技术,2008,31(11):814-819.
    [4]段晓礁,谭志新,兰小飞,等.用20—1020keV单能质子刻度CR-39固体核径迹探测器[J].物理学报,2010,59(5):3147-3153.
    [5]王芳,马新兴,史廷明,等.CR39中子个人监测系统的建立[J].中国辐射卫生,2013,22(2):213-214.
    [6]曹磊,邓君,王成国,等.用于油气田测井的CR39快/热中子个人剂量计的初步设计[J].中华放射医学与防护杂志,2011,31(4):489-492.
    [7]郝启辰,过惠平,王洪超.火花计数器应用于CR-39测氡计数研究[J].科学技术与工程,2012,12(36):9968-9971.
    [8]肖拥军,周青芝,汤凌志.湿度对CR-39固体核径迹探测器测量220Rn/222Rn水平的影响研究[J].南华大学学报(自然科学版),2010(2):15-17.
    [9]王光昶,张建炜,郑志坚.强激光质子束透视成像应用于肿瘤细胞异型性分析的探索研究[J].中国医学物理学杂志,2010,27(2):1777-1780.
    [10]唐秀欢,江新标,王鹏,等.亚细胞微观硼浓度测量方法研究[J].原子能科学技术,2008,42(6):517-521.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700