正常角膜伴近视散光眼角膜各形态参数分布特征及与全眼屈光参数的关系
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:Investigation of corneal shape parameters distribution in normal cornea with myopic astigmatism and correlation with ocular refractive parameters
  • 作者:李勇 ; 李晶 ; 魏升升 ; 万雅群 ; 叶璐 ; 李娟 ; 刘建国
  • 英文作者:LI Yong;LI Jing;WEI Sheng-Sheng;WAN Ya-Qun;YE Lu;LI Juan;LIU Jian-Guo;Refractive Surgery Center,Department of Ophthalmology,Xi ’an No. 4 Hospital;
  • 关键词:角膜形态 ; Pentacam眼前节分析 ; 全眼屈光参数 ; 角膜屈光手术 ; 近视散光
  • 英文关键词:corneal shape;;Pentacam scheimpflug system;;ocular refractive parameters;;corneal refractive surgery;;myopic astigmatism
  • 中文刊名:XKJZ
  • 英文刊名:Recent Advances in Ophthalmology
  • 机构:西安市第四医院眼科激光视力矫正中心;
  • 出版日期:2016-01-04 17:14
  • 出版单位:眼科新进展
  • 年:2016
  • 期:v.36;No.235
  • 基金:西安市社发-医学攻关项目[编号:SF1315(3)];; 陕西省卫生科研项目(编号:2014D49)~~
  • 语种:中文;
  • 页:XKJZ201601015
  • 页数:3
  • CN:01
  • ISSN:41-1105/R
  • 分类号:56-58
摘要
目的探讨正常角膜伴近视散光眼角膜各形态参数分布特征及与全眼屈光参数的关系。方法收集102例(204眼)近视患者,分别应用非接触眼压计、检影仪、A超、角膜超声测厚仪、Pentacam三维眼前节分析仪进行眼压、屈光度、眼轴长度、角膜厚度、角膜前后表面形态参数等测量,应用SPSS 13.0统计学软件进行数据分析。结果正常角膜伴近视散光眼角膜中央厚度(537.06±25.45)μm,角膜中央前表面高度(2.35±1.39)μm,中央后表面高度(3.60±2.73)μm,角膜Q值(8 mm范围)-0.30±0.10,角膜最薄点距角膜顶点距离(0.59±0.19)mm,角膜最薄点位置多位于角膜顶点颞下方,占91.67%。角膜最薄点至周边厚度增长变化指数为1.04±0.13。角膜中央前后表面高度、角膜Q值、角膜最薄点至周边厚度增长变化指数与等效球镜、眼压、眼轴长度无明显相关性(均为P>0.05)。结论正常角膜伴近视散光眼角膜形态各项参数分布值各有特点,与全眼屈光参数无明显相关性。
        Objective To investigate the distribution and characteristics of normal corneal shape parameters in myopic astigmatism as well as relationship with ocular refractive parameters.Methods A total of 102 myopic patients(204 eyes) with astigmatism were collected,and the intraocular pressure,refraction pcular axis length,corneal thickness,anterior and posterior corneal surface shape were measured with non-contact tonometer,mydriatic optometry,A-scan,Pentacam scheimpflug system and ultrasonic pachymeter,respectively.The data was analyzed by SPSS 13.0software.Results The corneal central thickness was(537.06 ±25.45) μm,the corneal central anterior surface elevation and posterior surface elevation were(2.35 ±1.39) μm and(3.60±2.73) μm.The mean corneal Q value with 8 mm was-0.30 ±0.10,the mean distance of corneal apex and corneal thinnest thickness was(0.59 ±0.19) mm,and the thinnest thickness mostly located in the temporal inferior of the apex mostly(91.67%).The Pachymetric progression index average was 1.04 ±0.13.There was no significant correlation between corneal anterior surface elevation,posterior surface elevation,Q value,pachymetric progression index average and spherical equivalent,intraocular pressure,ocular axis length,respectively(all P > 0.05).Conclusion Characteristic of corneal shape parameters distribution in normal cornea with myopic astigmatism are not identical.There was no significant difference between corneal shape parameters of myopic patients and total ocular refractive parameters.
引文
[1]AL-AGEEL S,AL-MUAMMAR AM.Comparison of central cornealthickness measurements by Pentacam,noncontact specular microscope,and ultrasound pachymetry in normal and post-LASIK eyes[J].Saudi J Ophthalmol,2009,23(3-4):181-187.
    [2]钱娜娜,余腾.Sirius与自动角膜曲率仪测量角膜曲率的一致性分析[J].新乡医学院学报,2015,32(8):744-748.
    [3]ISHIBAZAWA A,IGARASHI S,HANADA K,NAGAOKA T,ISHIKO S,ITO H,et al.Central corneal thickness measurements with Fourier-domain optical coherence tomography versus ultrasonic pachymetry and rotating Scheimpflug camera[J].Cornea,2011,30(6):615-619.
    [4]ORUCOGLU F,AKM AN M,ONAL S.Analysis of age,refractive error and gender related changes of the cornea and the anterior segment of the eye with Scheimpflug imaging[J].Cont Lens Anterior Eye,2015,4(21):S1367.
    [5]HAN-BOR F,KOOI-LIN G L.Corneal elevation indices in normal and keratoconic eyes[J].J Cataract Refract Surg,2006,32(8):1281-1287.
    [6]ORKUN M,ORHAN A,KEMAL O.Posterior corneal elevation and back difference corneal elevation in diagnosing forme fruste keratoconus in the fellow eyes of unilateral keratoconus patients[J].J Cataract Refract Surg,2013,39(9):1348-1357.
    [7]LUZ A,URSULIO M,CASTANEDA D.Corneal thickness progression from the thinnest point to the limbus:study based on a normal and a keratoconus population to create reference values[J].Arq Bras Oftalmol,2006,69(4):579-583.
    [8]王勤美.角膜断层地形图学[M].北京:人民卫生出版社,2014:51-63.
    [9]Holladay JT.Optic quality and refractive surgery[J].Int Ophthalmol Clin,2003,43(2):119-136.
    [10]刘俐利,陈辉.Pentacam眼前节测量分析系统对青壮年近视患者角膜前、后表面Q值的测量和分析[J].眼视光学杂志,2009,11(1):23-26.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700