摘要
使用岛津LC-30A超高效液相色谱仪建立了水体中阿特拉津的快速检测方法。采用C18色谱柱,甲醇/水=65/35(v/v)为流动相,紫外检测器的检测波长为223 nm,二氯甲烷作为萃取剂。在0.03~1.0μg/mL的浓度范围内标准品的线性相关系数为0.999 4,曲线线性关系良好。对添加浓度为1.0μg/L的6个平行样品进行分析,保留时间的RSD%为0.068%,浓度结果的RSD%为4.183%,结果的重现性良好,在此浓度下平均回收率为97%。研究结果显示,方法的最低检出限为0.1μg/L,最低定量限为0.35μg/L,具有较高的检测灵敏度。
A method was established for rapid determining of atrazine in water. The analysis was performed by a LC-30A UHPLC system of Shimadzu,using a C18 column for the stationary phase and methanol/water=65/35(v/v)as the mobile phase.The detection wavelength of UV detector was 223 nm. Dichloromethane was used as extractant. The calibration curve of atrazine showed good linearity in the range of 0.03~1.0 μg/m L,with excellent correlation coefficients which was 0.999 4. The RSD% of retention time and concentration results was 0.068% and4.183% respectively when analysis of six parallel samples with a concentration of 1.0 μg/L. The reproducibility of the results is good. The average recovery was 97% under this concentration. LOD and LOQ was 0.1 μg/L and 0.35μg/L respectively,which showed good sensitivity.
引文
[1]吴柳彦,刘笑寒,王梦乔,等.水体中阿特拉津去除技术研究进展[J].环境保护科学,2013(5):10-14.
[2]高远,杨帆,秦景.阿特拉津环境污染危害及防治对策[J].水利技术监督,2014(2):11-13.
[3]胡鸿雁,张红雨,张杰.固相萃取—高效液相色谱法测定水中阿特拉津[J].环境科学与技术,2006(12):31-32.
[4]国家环境保护部. HJ 587-2010水质阿特拉津的测定[S].2010.
[5]黄齐湘,成院飞.气相色谱—质谱法测定蔬菜中莠去津残留量[J].科研,2016,7(37):166-167.
[6]国家环境保护总局. GB 3838-2002地表水环境质量标准[S]. 2002.