离子色谱法测定家庭饮用水中阴阳离子
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:Determination of Anions and Cations in Household Drinking Water by Ion Chromatography
  • 作者:程琛 ; 刘佳俊 ; 林曼利 ; 唐银凤
  • 英文作者:CHENG Chen;LIU Jia-jun;LIN Man-li;TANG Yin-feng;School of Resources and Civil Engineering,Suzhou University;China Guodian Chifeng Chemical Co.,Ltd;
  • 关键词:离子色谱法 ; 阴离子 ; 阳离子 ; 钙镁硬度
  • 英文关键词:ion chromatography;;anion;;cation;;calcium and magnesium hardness
  • 中文刊名:ZJKN
  • 英文刊名:Journal of Hebei North University(Natural Science Edition)
  • 机构:宿州学院资源与土木工程学院;中国国电赤峰化工有限公司;
  • 出版日期:2018-05-28
  • 出版单位:河北北方学院学报(自然科学版)
  • 年:2018
  • 期:v.34;No.145
  • 基金:安徽省教育厅自然科学研究重点项目(KJ2017A445);; 宿州学院重点科研项目(2016yzd08);; 宿州区域发展协同创新项目(2015SZXTXSKF07)
  • 语种:中文;
  • 页:ZJKN201805002
  • 页数:6
  • CN:05
  • ISSN:13-1360/N
  • 分类号:6-11
摘要
目的建立离子色谱法测定家庭饮用水中常规阴阳离子的方法,为居民饮用水安全提供可靠数据。方法采用ICS-900型离子色谱仪测试水样中F~-、Cl~-、NO_3~-、SO_4~(2-)、PO_4~(3-)、Li~+、Na~+、NH_4~+、K~+、Mg~(2+) 和 Ca~(2+)共11种常规阴阳离子含量,测试时选择合适的仪器条件,对11份水样进行测定,阴离子分析条件为KOH淋洗液浓度10~40mmol·L~(-1)、抑制器电流99mA,阳离子仪器条件为甲烷磺酸淋洗液20mmol·L~(-1)、抑制器电流75mA,淋洗液流速均为1.0mL·min~(-1),进样量均为10μL。结果 3次测量取平均值且相对标准偏差均小于1%,表明离子色谱法测定水溶液中各阴阳离子准确、高效、误差小、重复性好。阴阳离子色谱柱在该试验条件下对各离子的分离效果较好,标准曲线相关性均在0.999以上。结论家庭生活中自来水水样中Mg~(2+)和Ca~(2+)含量较高,水体硬度较大,但符合生活饮用水水质标准。
        Objective To provide reliable data for drinking water safety,method for the determination of anions and cations in drinking water by ion chromatography was established.Methods The 11 kinds of conventional antion and cation F~-,Cl~-,NO_3~-,SO_4~(2-),PO_4~(3-),Li~+,Na~+,NH_4~+,K~+,Mg~(2+) and Ca~(2+) were monitored by the ICS-900 ion chromatography;the appropriate testing conditions of instrument were chosen;The anion analysis conditions for KOH eluent concentration was 10 mmol/L~40 mmol/L,and the suppressor current was 99 mA;the cation instrument conditions for the methane sulfonic acid eluent were 20 mmol/L,and 75 mA suppressor current;and the both eluent current speeds were 1.0 mL/min,and the sample quantity 10μL.Results The average value of three measurements and the relative standard deviations were less than 1%.The results show that the ion chromatography method on the determination of antions and cations in aqueous solution was accurate and high-efficient with small-error and good repeatability.Under the above experimental conditions,the separation effect to each ion of ion chromatographic column was good,and the correlation coefficient of the standard curve was above 0.999.Conclusion The results indicated that the concentrations of Mg~(2+) and Ca~(2+) were higher in household tap water and the water hardness was higher,but the water quality was in accordance with the standard of drinking water quality.
引文
[1]邵全忠,任锋,杨丹,等.周口市农村饮用水氟含量调查研究[J].地下水,2011,33(05):31-34.
    [2]栾奇.南阳市地下水污染健康风险评价[D].长春:吉林大学,2015.
    [3]张华妮.微量元素在原发性高血压中的作用[J].国外医学(医学地理分册),2012,33(04):288-289.
    [4]霍晶.离子色谱法测定水中7种阴离子[J].中国无机分析化学,2014,4(02):8-10.
    [5]黄文锋,张骊,肖琳.离子色谱法测定饮用水中4种阴离子与传统化学法的比较[J].中国卫生检验杂志,2013,23(02):309-312.
    [6]付志军,罗桂娟,李雅妍,等.离子选择电极法测定水中氯化物[J].环境监测管理与技术,2014,26(01):49-52.
    [7]范中学,杨晓栋,周蓉,等.2014年陕西省某县饮水型氟中毒改水效果评价[J].职业与健康,2016,32(02):231-234.
    [8]阳艳,骆灵喜,冯家望.利用离子色谱法与火焰原子吸收法测定水中钾钠的比较研究[J].广东化工,2015,42(13):261-263.
    [9]樊颖果,徐国津.原子吸收光谱和原子发射光谱法测定酸雨中钾、钠、钙、镁方法比较[J].中国无机分析化学,2013,3(02):28-31.
    [10]余秀娟,曾钰,那晶晶.离子色谱法同时测定水中8种阴离子[J].环保科技,2015,21(04):57-59.
    [11]李安军,汤有宏,刘国英,等.离子色谱法同时测定酿酒用水中6种阳离子的研究[J].酿酒,2015,42(05):19-21.
    [12]谢静,杨树科,戴建辉,等.离子色谱法测定瓶装水中阴离子[J].理化检验(化学分册),2011,47(07):808-810.
    [13]周贻兵,刘利亚,李磊,等.离子色谱法测定生活饮用水中F-、Cl-、NO3-、Na+、SO42-的含量[J].微量元素与健康研究,2014,31(04):53-54.
    [14]闫春辉,徐凌云,钱文虎,等.离子色谱法测定不同地区家庭饮用井水中的阴、阳离子[J].理化检验(化学分册),2015,51(02):211-214.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700