高分辨率3D CE-MRV联合3D T1增强在SEEG手术计划中的应用
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:Value of high-resolution three-dimensional contrast-enhanced magnetic resonance venography combined with three-dimensional T1-weighted magnetic resonance imaging in surgical planning for stereotactic electroencephalography
  • 作者:高薇 ; 毕永峰 ; 叶明 ; 吴超 ; 王中 ; 虞正权 ; 陈罡
  • 英文作者:Gao Wei;Bi Yongfeng;Ye Ming;Wu Chao;Wang Zhong;Yu Zhengquan;Chen Gang;Neurosurgery Department,The First Affiliated Hospital of Soochow University;
  • 关键词:立体定向脑电图 ; 对比增强磁共振静脉成像 ; 对比增强磁共振成像 ; 并发症
  • 英文关键词:stereotactic electroencephalography;;contrast-enhanced magnetic resonance venography;;T1-weighted magnetic resonance imaging;;complication
  • 中文刊名:ZQYK
  • 英文刊名:Journal of Chongqing Medical University
  • 机构:苏州大学附属第一医院神经外科;
  • 出版日期:2018-09-29 15:54
  • 出版单位:重庆医科大学学报
  • 年:2019
  • 期:v.44
  • 基金:国家自然科学基金资助项目(编号:81771255)
  • 语种:中文;
  • 页:ZQYK201901019
  • 页数:4
  • CN:01
  • ISSN:50-1046/R
  • 分类号:91-94
摘要
目的:探讨利用高分辨率3D CE-MRV联合3D T1增强在SEEG手术计划的可行性及准确性。方法:对计划行SEEG手术的12例药物难治性癫痫患者进行定制的MRI检查,利用iplan软件制定手术计划,共植入89根电极,分析手术结果。结果:12例患者共89根电极均顺利植入,术后未发生出血。其中包括2例岛叶癫痫的患者。结论:高分辨率3D CE-MRV有助于寻找无血管通道,3D T1增强有助于了解针道所经过的脑组织结构,两者联合使用可以准确地制定手术计划,减少手术并发症。
        Objective:To investigate the feasibility and accuracy of high-resolution three-dimensional contrast-enhanced magnetic resonance venography(3D CE-MRV)combined with three-dimensional T1-weighted magnetic resonance imaging(3D T1-weight MRI)in surgical planning for stereotactic electroencephalography. Methods:A total of 12 patients with medically intractable epilepsy who planned to undergo SEEG surgery were enrolled. Individualized MRI was performed,and the i Plan software was used to develop surgical plans. A total of 89 electrodes were implanted,and the results of the surgery were analyzed. Results:The 89 electrodes were successfully implanted,and no postoperative bleeding was observed. Of all patients,2 had insular epilepsy. Conclusion:High-resolution 3D CE-MRV helps to find the avascular trajectory and 3D T1-weight MRI helps to know the structure of brain tissue in the avascular trajectory. Their combination can accurately develop the surgical plan and reduce complications.
引文
[1]Cardinale F,Casaceli G,Raneri F,et al.Implantation of stereoelectroencephalography electrodes:a systematic review[J].J Clin Neurophysiol,2016,33(6):490-502.
    [2]Mullin JP,Shriver M,Alomar S,et al.Is SEEG safe?A systematic review and meta-analysis of stereo-electroencephalography-related complications[J].Epilepsia,2016,57(3):386-401.
    [3]Mirzayan MJ,von Roden M,Bulacio J,et al.The usefulness of intraoperative cerebral C-Arm CT angiogram for implantation of intracranial depth electrodes in stereotactic electroencephalography procedure[J].Stereotact Funct Neurosurg,2016,94(1):10-17.
    [4]Cardinale F,Cossu M,Castana L,et al.Stereoelectroencephalography:surgical methodology,safety,and stereotactic application accuracy in 500 procedures[J].Neurosurg,2013,72(3):353-366.
    [5]刘烁,林光耀,蒋杰,等.颅内静脉窦2D-TOF MR和3D-CE MRA扫描方法对比研究[J].磁共振成像,2015,6(3):178-181.
    [6]潘璟琍,张道春,丁建荣.高分辨CE-MRV对颅内静脉系统的显示[J].医学影像学杂志,2011,21(2):175-177.
    [7]Liu W,Guo H,Du X,et al.Cortical vessel imaging and visualization for image guided depth electrode insertion[J].Comput Med Imaging Graph,2013,37(2):123-130.
    [8]Cardinale F,Pero G,Quilici L,et al.Cerebral angiography for multimodal surgical planning in epilepsy surgery:description of a new three-dimensional technique and literature review[J].World Neurosurg,2015,84(2):358-367.
    [9]Yigˇit H,Turan A,Ergün E,et al.Time-resolved MR angiography of the intracranial venous system:an alternative MR venography technique[J].Eur Radiol,2012,22(5):980-989.
    [10]Minkin K,Gabrovski K,Penkov M,et al.Stereoelectroencephalography using magnetic resonance angiography for avascular trajectory planning:technical report[J].Neurosurgery,2017,81(4):688-695.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700