Comparative performance assessment of commercially available automatic external defibrillators: A simulation and real-life measurement study of hands-off time
详细信息    查看全文
文摘
Early and good quality cardiopulmonary resuscitation (CPR) and the use of automated external defibrillators (AEDs) improve cardiac arrest patients’ survival. However, AED peri- and post-shock/analysis pauses may reduce CPR effectiveness.MethodsThe time performance of 12 different commercially available AEDs was tested in a manikin based scenario; then the AEDs recordings from the same tested models following the clinical use both in Pavia and Ticino were analyzed to evaluate the post-shock and post-analysis time.ResultsNone of the AEDs was able to complete the analysis and to charge the capacitors in less than 10 s and the mean post-shock pause was 6.7 ± 2.4 s. For non-shockable rhythms, the mean analysis time was 10.3 ± 2 s and the mean post-analysis time was 6.2 ± 2.2 s. We analyzed 154 AED records [104 by Emergency Medical Service (EMS) rescuers; 50 by lay rescuers]. EMS rescuers were faster in resuming CPR than lay rescuers [5.3 s (95%CI 5–5.7) vs 8.6 s (95%CI 7.3–10).ConclusionsAEDs showed different performances that may reduce CPR quality mostly for those rescuers following AED instructions. Both technological improvements and better lay rescuers training might be needed.
NGLC 2004-2010.National Geological Library of China All Rights Reserved.
Add:29 Xueyuan Rd,Haidian District,Beijing,PRC. Mail Add: 8324 mailbox 100083
For exchange or info please contact us via email.