Near and intermediate visual and reading performance of patients with a multifocal apodized diffractive intraocular lens using an electronic reading desk
详细信息    查看全文
文摘
To functionally evaluate an apodized diffractive multifocal intraocular lens (IOL) using an electronic reading desk (Salzburg Reading Desk).

Setting

University Eye Clinic Heidelberg, Heidelberg, Germany.

Design

Prospective case series.

Methods

After implantation of an Acrysof IQ Restor +3.0 diopter (D) multifocal IOL, uncorrected (UDVA) and corrected (CDVA) distance visual acuities, uncorrected (UNVA) and distance-corrected (DCNVA) near visual acuities, and uncorrected (UIVA) and distance-corrected (DCIVA) intermediate visual acuities were determined. Criteria for evaluating reading acuity were the reading speed and print size. Monocular and binocular examinations were performed with the electronic reading desk at fixed distances and at the patient's preferred near and intermediate distances. In addition, a subjective questionnaire was administered.

Results

Forty eyes (20 patients) were enrolled. Monocularly, the medians were UDVA, 0.00 logMAR (range 0.26 to −0.14 logMAR); CDVA, −0.08 logMAR (range 0.16 to −0.24 logMAR); 40 cm UNVA, 0.04 logMAR (range 0.24 to −0.10 logMAR); DCNVA, 0.00 logMAR (range 0.18 to −0.22 logMAR); 80 cm UIVA, 0.15 logMAR (range 0.40 to −0.18 logMAR); DCIVA, 0.16 logMAR (range 0.42 to −0.14 logMAR). The monocular reading desk examinations resulted in an uncorrected reading acuity of 0.11 logMAR (range 0.49 to −0.01 logMAR) at 40 cm and 0.30 logMAR (range 0.48 to 0.00 logMAR) at 80 cm. Patients reported a high rate of spectacle independence and satisfaction in everyday life.

Conclusion

The +3.0 D IOL provided good postoperative functional results at far and near and improved the visual and reading acuity at an intermediate distance.