Impact of INTERMACS Profile on Clinical Outcomes for Patients Supported With the Total Artificial Heart
文摘
Insufficient data delineate outcomes for Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) profile 1 patients with the total artificial heart (TAH).

Methods

We studied 66 consecutive patients implanted with the TAH at our institution from 2006 through 2012 and compared outcome by INTERMACS profile. INTERMACS profiles were adjudicated retrospectively by a reviewer blinded to clinical outcomes.

Results

Survival after TAH implantation at 6 and 12 months was 76% and 71%, respectively. INTERMACS profile 1 patients had decreased 6-month survival on the device compared with those in profiles 2–4 (74% vs 95%, log rank: P = .015). For the 50 patients surviving to heart transplantation, the 1-year posttransplant survival was 82%. There was no difference in 1-year survival when comparing patients in the INTERMACS 1 profile with less severe profiles (79% vs 84%; log rank test P = .7; hazard ratio [confidence interval] 1.3 [0.3–4.8]).

Conclusions

Patients implanted with the TAH as INTERMACS profile 1 had reduced survival to transplantation compared with less sick profiles. INTERMACS profile at the time of TAH implantation did not affect 1-year survival after heart transplantation.