A Novel Thermal-activated Shape Memory Penile Prosthesis: Comparative Mechanical Testing
详细信息    查看全文
文摘
To compare a novel nickel-titanium (Ni-Ti) shape memory alloy (SMA) penile prosthesis of our own design with commercially available prostheses using a format similar to mechanical testing done at major penile prosthesis manufacturers. We evaluated the mechanical parameters of commercially available penile prostheses and used this information to guide the development of the Ni-Ti-based physiological penile prosthesis that expands and becomes erect with a small amount of heat applied.MethodsA penile prosthesis consisting of an exoskeleton of temperature-tuned Nitinol was designed and prototyped. Mechanical testing was performed in a model of penile buckling, penile lateral deviation, and original penile shape recovery commonly used by penile prosthesis manufacturers for testing.ResultsOur SMA penile prosthesis demonstrated useful mechanical characteristics, including rigidity to buckling when activated similar to an inflatable penile prosthesis (2.62 kgf SMA vs 1.42 kgf inflatable penile prosthesis vs 6.45 kgf for a malleable prosthesis). The Ni-Ti also became more pliable when deactivated within acceptable mechanical ranges of existing devices. It could be repeatedly cycled and generate a restorative force to become erect.ConclusionAn SMA-based penile prosthesis represents a promising new technology in the treatment of erectile dysfunction. We demonstrated that an Ni-Ti-based prosthesis can produce the mechanical forces necessary for producing a simulated erection without the need for a pump or reservoir, comparable with existing prostheses.
NGLC 2004-2010.National Geological Library of China All Rights Reserved.
Add:29 Xueyuan Rd,Haidian District,Beijing,PRC. Mail Add: 8324 mailbox 100083
For exchange or info please contact us via email.