Location dependency of the spatial resolution of cone beam computed tomography for dental use
详细信息    查看全文
文摘
| Figures/TablesFigures/Tables | ReferencesReferences

Objective

This study evaluates the spatial resolution of cone beam computed tomography for dental use (CBCT) via modulation transfer function (MTF) analyses.

Study Design

Two models of CBCT system, 3DX FPD8 and FineCube v.12, were used. MTF analysis was applied to each CBCT system using a thin tungsten wire technique.

Results

The MTF curves in the radial direction on the XY-plane were concordant regardless of position, whereas the curves in the azimuthal direction tended to decrease as the distance from the rotation center increased. In the Z-axis direction, the MTF curve of the medial level of the field of view was superior to that of any other level.

Conclusion

The spatial resolution of CBCT systems depends on the location within the field of view. Because the spatial resolution was the highest in the medial level and rotation center position, an object should be placed at this position during a CBCT examination.