Explaining and reducing the variation in inter-laboratory reported values for International Normalised Ratio
详细信息    查看全文
文摘
Significant discrepancies identified in reported INR values vs those generated using their values for PT, ISI, and MNPT. Discrepancies also identified producing significant variation in ISI, MNPT and INR for same reagent lot/instrument combinations. These data partially explain the large inter-laboratory variation in test results reported for INR. Pre- and/or post-analytical (rather than analytical) issues may thus be the main driver for variation in laboratory reported INR values.