Accuracy of user-adjusted axial length measurements with optical biometry in eyes having combined phacovitrectomy for macular-off rhegmatogenous retinal detachment
详细信息    查看全文
文摘
To evaluate the accuracy of user-adjusted axial length (AL) measured by optical biometry for intraocular lens (IOL) calculations in eyes having combined phacovitrectomy for macula-off rhegmatogenous retinal detachment (RRD).

Setting

Ophthalmology Department, Calderdale Royal Hospital, Halifax, United Kingdom.

Design

Retrospective case series.

Methods

Consecutive eyes having phacovitrectomy for macula-off RRD were evaluated. The ALs were measured using optical biometry with user adjustment to identify a posterior peak corresponding to the eye's AL and ultrasound (US). These were compared with each other and with the postoperative optical biometry and analyzed for accuracy as an indication of the eye's AL.

Results

The study comprised 22 eyes of 22 patients. There was no statistically significant difference between the mean AL measurements derived from user-adjusted optical biometry and US AL (P = .964). The user-adjusted optical biometry was not statistically significantly different from the postoperative optical biometry (P = .242). Compared with the postoperative optical biometry, the IOL power was within ±0.5 diopter in 12 (92%) of 13 cases (95% confidence interval [CI], 77.8 to 100.0) for user-adjusted optical biometry and in 10 (77%) of 13 cases (95% CI, 54.0 to 99.8) for US measurements.

Conclusions

User-adjusted optical biometry could be used as an alternative method for measuring AL in macula-off RRD with combined phacovitrectomy. However, optical biometry would require assessment of agreement with US AL in cases in which a posterior peak is not easily identifiable. User-adjusted optical biometry might outperform US AL when calculating IOL power; however, a larger study should be performed to confirm this.