Automated Breast Ultrasound: Dual-Sided Compared with Single-Sided Imaging
文摘
The design and performance of a mammographically configured, dual-sided, automated breast ultrasound (ABUS) 3-D imaging system are described. Dual-sided imaging (superior and inferior) is compared with single-sided imaging to aid decisions on clinical implementation of the more complex, but potentially higher-quality dual-sided imaging. Marked improvement in image quality and coverage of the breast is obtained in dual-sided ultrasound over single-sided ultrasound. Among hypo-echoic masses imaged, there are increases in the mean contrast-to-noise ratio of 57% and 79%, respectively, for spliced dual-sided versus superior or inferior single-sided imaging. The fractional breast volume coverage, defined as the percentage volume in the transducer field of view that is imaged with clinically acceptable quality, is improved from 59% in both superior and inferior single-sided imaging to 89% in dual-sided imaging. Applying acoustic coupling to the breast requires more effort or sophisticated methods in dual-sided imaging than in single-sided imaging.