Isomer-Selective Detection of Aromatic Molecules in Temperature-Programmed Desorption for Model Catalysis
详细信息    查看全文
文摘
Based on three different molecules dosed on a Pt(111) single crystal the selectivity and sensitivity of REMPI-TPD in UHV is investigated for a potential application in heterogeneous catalysis. It is shown that the two structural isomers ethylbenzene and p-xylene can be discriminated by REMPI in a standard TPD experiment. The latter is not possible for the ionization with electrons in a Q-MS. It is further demonstrated by benzene TPD studies that the sensitivity of the REMPI-TOF-MS is comparable to commercial EI-Q-MS solutions and enables the detection of less than 0.6% molecules of a monolayer.