Closure of Interatrial Defects by Immediate-Release Patch
详细信息    查看全文
  • 作者:Aliakbar Zeinaloo (1)
    Keyhan Sayadpour Zanjani (1)
    Bahman Rastkar (2)
    Alimohammad Hajizeinali (3)
    Bahram Sohrabi (2)
    Mahmoud Samadi (2)
    Eleftherios B. Sideris (4)
  • 关键词:Fenestration ; Immediate ; Release Patch ; Total cavopulmonary connection ; Transcatheter Patch
  • 刊名:Pediatric Cardiology
  • 出版年:2012
  • 出版时间:December 2012
  • 年:2012
  • 卷:33
  • 期:8
  • 页码:1253-1258
  • 全文大小:320KB
  • 参考文献:1. Sideris EB, Macuil B, Sideris VE, Moulopoulos SD (2005) Abstracts of the 54th Annual Scientific Session of the American College of Cardiology, 6- March 2005, Orlando. J Am Coll Cardiol 45:1A-46A
    2. Calachanis M, Macuil B, Zamora R, Coulson J, Toumanides S, Sideris E (2007) Effectiveness of transcatheter patch release with surgical adhesives in the occlusion of heart defects. Abstracts of the 42nd Annual Meeting of the Association for European Paediatric Cardiology, 16-9 May 2007, Warsaw, Poland. Cardiol Young 17(Suppl 1):1-9
    3. Amin Z, Hijazi ZM, Bass JL, Cheatham JP, Hellenbrand WE, Kleinman CS (2004) Erosion of Amplatzer septal occluder device after closure of secundum atrial septal defects: review of registry of complications and recommendations to minimize future risk. Catheter Cardiovasc Interv 63:496-02 CrossRef
    4. DiBardino DJ, McElhinney DB, Kaza AK, Mayer JE Jr (2009) Analysis of the US food and drug administration manufacturer and user facility device experience database for adverse events involving Amplatzer septal occluder devices and comparison with the Society of Thoracic Surgery congenital cardiac surgery database. J Thorac Cardiovasc Surg 137:1334-341 CrossRef
    5. Divekar A, Gaamangwe T, Shaikh N, Raabe M, Ducas J (2005) Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder. J Am Coll Cardiol 45:1213-218 CrossRef
    6. El-Said HG, Moore JW (2009) Erosion by the Amplatzer septal occluder: experienced operator opinions at odds with manufacturer recommendations? Catheter Cardiovasc Interv 73:925-30 CrossRef
    7. Khodaverdian RA, Jones KW (2009) Metal allergy to Amplatzer occluder device presented as severe bronchospasm. Ann Thorac Surg 88:2021-022 CrossRef
    8. Lai DW, Saver JL, Araujo JA, Reidl M, Tobis J (2005) Pericarditis associated with nickel hypersensitivity to the Amplatzer occluder device: a case report. Catheter Cardiovasc Interv 66:424-26 CrossRef
    9. Levi DS, Moore JW (2004) Embolization and retrieval of the Amplatzer septal occluder. Catheter Cardiovasc Interv 61:543-47 CrossRef
    10. Marini D, Boudjemline Y, Agnoletti G (2007) Closure of extracardiac Fontan fenestration by using the covered Cheatham platinum stent. Catheter Cardiovasc Interv 69:1002-006 CrossRef
    11. Meraji SM, Ghaffari S, Zanjani KS (2006) Occlusion of interatrial fenestration with the Amplatzer septal occluder device: first case report from Iran. Iranian Heart J 7:56-9
    12. Mullins CE (2002) Frameless patch occlusion of atrial septal defects: frameless–less secure. Catheter Cardiovasc Interv 57:408-09 CrossRef
    13. Rezaian GR, Amirghofran AA, Afifi S, Moaref AR, Rezaian S (2011) Nitinol wire mesh fracture and traumatic left atrial thrombus in a patient with atrial septal defect amplatzer occluder. J Card Surg 26:41-3 CrossRef
    14. Sideris EB, Moulopoulos SD (2004) Transcatheter patch occlusion of heart defects. Catheter Cardiovasc Interv 61:150-51 CrossRef
    15. Sideris EB, Kaneva A, Sideris SE, Moulopoulos SD (2000) Transcatheter atrial septal defect occlusion in piglets by balloon detachable devices. Catheter Cardiovasc Interv 51:529-34 CrossRef
    16. Sideris EB, Toumanides S, Macuil B, Gutierrez-Leonard H, Poursanov M, Sokolov A, Moulopoulos SD (2002) Transcatheter patch correction of secundum atrial septal defects. Am J Cardiol 89:1082-086 CrossRef
    17. Sideris EB, Sideris CE, Stamatelopoulos SF, Moulopoulos SD (2002) Transcatheter patch occlusion of experimental atrial septal defects. Catheter Cardiovasc Interv 57:404-07 CrossRef
    18. Sideris E, Chamie F, Haddad G, Calachanis M, Toumanides S (2008) Transcatheter patch correction of nonsecundum atrial septal defects: patient selection and clinical results. Cardiol Young 18:1-06
    19. Sideris B, Sideris E, Calachanis M, Papantoniou V, Moulopoulos S (2010) The Immediate-Release Patch in the correction of experimental atrial septal defects. Catheter Cardiovasc Interv 76:572-77 CrossRef
    20. Spence MS, Qureshi SA (2005) Complications of transcatheter closure of atrial septal defects. Heart 91:1512-514 CrossRef
    21. Taggart NW, Dearani JA, Hagler DJ (2011) Late erosion of an Amplatzer septal occluder device 6?years after placement. J Thorac Cardiovasc Surg 142:221-22 CrossRef
    22. Tehrani RB, Rostami A, Basiri HA, Mortezaiian H (2008) Device-induced perforation of right atrium following interventional closure of atrial septal defect (ASD). J Teh Univ Heart Ctr 3:113-15
    23. Wertman B, Azarbal B, Riedl M, Tobis J (2006) Adverse events associated with nickel allergy in patients undergoing percutaneous atrial septal defect or patent foramen ovale closure. J Am Coll Cardiol 47:1226-227 CrossRef
    24. Wilson WM, Goh TH, Oqueli E, Willis Y, Chan C, Skillington P, Grigg LE (2010) Aorto-left atrial fistula post-percutaneous device ASD closure. Heart Lung Circ 19:53-5 CrossRef
    25. Zeinaloo A, Meraji SM, Zanjani KS, Mirzaaghayan MR (2008) Device occlusion versus surgery for closure of congenital heart defects: cost issues in Iran. J Teh Univ Heart Ctr 3:141-44
    26. Zeinaloo A, Macuil B, Zanjani KS, Sideris E (2011) Transcatheter patch occlusion of ventricular septal defect in Down syndrome. Am J Cardiol 107:1838-840 CrossRef
  • 作者单位:Aliakbar Zeinaloo (1)
    Keyhan Sayadpour Zanjani (1)
    Bahman Rastkar (2)
    Alimohammad Hajizeinali (3)
    Bahram Sohrabi (2)
    Mahmoud Samadi (2)
    Eleftherios B. Sideris (4)

    1. Children’s Medical Center, Tehran University of Medical Sciences, 62 Dr Gharib Street, 14194, Tehran, Iran
    2. Shahid Madani Heart Hospital, Tabriz University of Medical Sciences, Tabriz, Iran
    3. Tehran Heart Center, Tehran University of Medical Sciences, Tehran, Iran
    4. Athenian Institute of Pediatric Cardiology, Athens, Greece
  • ISSN:1432-1971
文摘
The Immediate-Release Patch is the newest version of the Transcatheter Patch, which is a bioabsorbable device for the closure of cardiac defects. Closure of 12 atrial septal defects and 1 fenestration was attempted using this device. Of the 13 devices used, 12 were implanted successfully (including that for the fenestration), whereas 1 device moved from the original position and was retrieved percutaneously. No other major adverse events occurred. After a median follow-up period of 11?months, trivial (?-mm) residual shunts remained in two patients, and a significant residual defect (7?mm) remained in 1 patient. The major advantages of this device include its wirelessness, its bioabsorbablity, potential application for defects up to 30?mm using only three sizes, its easier retrieval, and its ability to occlude defects with a deficient rim and some ostium primum and sinus venosus type defects. In contrast to the metal devices, it is bulkier and has a different application method requiring operator familiarity, and it sometimes leaves residual shunts.
NGLC 2004-2010.National Geological Library of China All Rights Reserved.
Add:29 Xueyuan Rd,Haidian District,Beijing,PRC. Mail Add: 8324 mailbox 100083
For exchange or info please contact us via email.