Stability evaluation and correction of a pulsed neutron generator prompt gamma activation analysis system
- 作者:Roman J. Shypailo
- 关键词:Body composition ; Neutron activation analysis ; Nitrogen ; Humans ; Neutron generator
- 刊名:Journal of Radioanalytical and Nuclear Chemistry
- 出版年:2016
- 出版时间:March 2016
- 年:2016
- 卷:307
- 期:3
- 页码:1781-1786
- 全文大小:636 KB
- 参考文献:1.Barzilov AP, Novikov IS, Womble PC (2012) Gamma radiation. In: Adrovic F (ed.). InTech Open Access Publisher [eBooks]. http://www.healthwise.org . Accessed 24 Feb 2014
2.Nagai T, Fujii I, Muto H, Inouye T (1969) Total-body nitrogen and protein determined by in vivo fast-neutron activation analysis. J Nucl Med 10:192–196 3.Gierlik M, Batsch T, Moszynski M, Szczesniak T, Wolski D, Klamra W, Perot B, Perret G (2006) Comparative study of large NaI(Tl) and BGO scintillators for the EURopean illicit TRAfficking countermeasures kit project. IEEE Trans Nucl Sci 53:1737–1743CrossRef 4.Gordon CM, Peters CW (1990) A fast-neutron probe for tomography and bulk analysis. Intern J Appl Radiat Isot 41:1111–1116CrossRef 5.Vourvopoulos G, Womble PC (2001) Pulsed fast/thermal neutron analysis: a technique for explosives detection. Talanta 54(3):459–468CrossRef 6.Litz M, Waits C, Mullins J (2012) Neutron-activated gamma-emission: technology review, ARL-TR-5871. Army Research Lab, Adelphi 7.Shypailo RJ, Workeneh B, Ellis KJ (2013) Assessment of the associated particle prompt gamma neutron activation technique for total body nitrogen measurement in vivo. J Radioanal Nucl Chem 296:143–148 8.Beddoe AH, Zuidmeer A, Hill GL (1984) A prompt gamma in vivo neutron activation analysis facility for measurement of total body nitrogen in the critically ill. Phys Med Biol 29:371–383CrossRef 9.Chichester DL, Simpson JD, Lemchak M (2007) Advanced compact accelerator neutron generator technology for active neutron interrogation field work. J Radioanal Nucl Chem 271:629–637CrossRef 10.Reijonen J, Gicquel F, Hahto SK, King M, Lou TP, Leung KN (2005) D-D neutron generator development at LBNL. Appl Radiat Isot 63:757–763CrossRef 11.Fantidis JG, Nicolaou GE, Potolias C, Vordos N, Bandekas DV (2011) The comparison of four neutron sources for Prompt Gamma Neutron Activation Analysis (PGNAA) in vivo detections of boron. J Radioanal Nucl Chem 290:289–295CrossRef 12.Danon Y (2012) Pyroelectric crystal D-D and D-T neutron generators. J Instrum 7(4):C04002. doi:10.1088/1748-0221/7/04/C04002 CrossRef 13.Gillich D, Kovanen A, Herman B, Fullem T, Danon Y (2009) Pyroelectric crystal neutron production in a portable prototype vacuum system. Nucl Instrum Methods Phys Res Sect A 602:306–310CrossRef 14.Kramer GH, Burns LC, Noel L (1991) The BRMD BOMAB phantom family. Health Phys 61:895–902CrossRef 15.Russo P, Vo D (2005) Gamma-ray detectors for nondestructive analysis. Los Alamos National Laboratory. Report LA-UR-05-3813
- 作者单位:Roman J. Shypailo (1)
1. Department of Pediatrics, USDA/ARS Children’s Nutrition Research Center, Baylor College of Medicine, 1100 Bates Street, Houston, TX, 77030, USA
- 刊物类别:Chemistry and Materials Science
- 刊物主题:Chemistry
Nuclear Chemistry Physical Chemistry Nuclear Physics, Heavy Ions and Hadrons Diagnostic Radiology Inorganic Chemistry
- 出版者:Akad茅miai Kiad贸, co-published with Springer Science+Business Media B.V., Formerly Kluwer Academic
- ISSN:1588-2780
文摘
Source output stability is important for accurate measurement in prompt gamma neutron activation. This is especially true when measuring low-concentration elements such as in vivo nitrogen (~2.5 % of body weight). We evaluated the stability of the compact DT neutron generator within an in vivo nitrogen measurement system. Review of gamma event/time patterns and data from an auxiliary detector showed significant variations among repeated phantom runs. Neutron generator instability had a significant effect on measurement precision. The neutron generator used in our system must be monitored for output consistency. Adjustments must be made to measurement results to correct for generator instability.
| |