Calibration of Radiation Thermometers up to \(3000\,^{\circ }\hbox {C}\) : Eff
详细信息    查看全文
  • 作者:O. Kozlova ; S. Briaudeau ; L. Rongione ; F. Bourson…
  • 关键词:Blackbody source ; Emissivity ; High temperature ; Radiation thermometer calibration technique ; Thermal gradient
  • 刊名:International Journal of Thermophysics
  • 出版年:2015
  • 出版时间:August 2015
  • 年:2015
  • 卷:36
  • 期:8
  • 页码:1726-1742
  • 全文大小:1,303 KB
  • 参考文献:1.E.R. Woolliams, G. Machin, D.H. Lowe, R. Winkler, Metrologia 43, R11 (2006)CrossRef ADS
    2.H. Preston-Thomas, Metrologia 77, 3 (1990)CrossRef ADS
    3.Linearpyrometer LP3 Operating Instructions, August 2002, 5-TN-1622-02 (2002)
    4.M. Sadli, K. Anhalt, F. Bourson, S. Schiller, J. Hartmann, Int. J. Thermophys. 30, 69 (2009)CrossRef ADS
    5.K. Chahine, M. Ballico, J. Reizes, J. Madadnia, Int. J. Thermophys. 29, 386 (2008)CrossRef ADS
    6.A.V. Prokhorov, Metrologia 35, 465 (1998)CrossRef ADS
    7.S. Galal Yousef, P. Sperfeld, J. Metzdorf, Metrologia 37, 365 (2000)CrossRef ADS
    8.L.M. Hanssen, S.N. Mekhontsev, J. Zeng, V. Prokhorov, Int. J. Thermophys. 29, 352 (2008)CrossRef ADS
    9.H.J. Patrick, L.M. Hanssen, J. Zeng, T.A. Germer, Metrologia 49, S81 (2012)CrossRef ADS
    10.A.V. Kostanovskii, M.G. Zeodinov, M.E. Kostanovskaya, High Temp. 43, 793 (2005)CrossRef
    11.T.P. Jones, J. Tapping, Metrologia 18, 23 (1982)CrossRef ADS
    12.P. Saunders, in Proceedings of Ninth International Temperature Symposium (Los Angeles), Temperature: Its Measurement and Control, in Science and Industry, ed. by C.W. Meyer, AIP Conference Proceedings 1552, vol. 8 (AIP, Melville, NY, 2013), p. 619
  • 作者单位:O. Kozlova (1)
    S. Briaudeau (1)
    L. Rongione (1)
    F. Bourson (1)
    S. Guimier (1)
    S. Kosmalski (1)
    M. Sadli (1)

    1. Laboratoire Commun de Métrologie LNE-CNAM, La Plaine Saint-Denis, France
  • 刊物类别:Physics and Astronomy
  • 刊物主题:Physics
    Condensed Matter
    Mechanics
    Industrial Chemistry and Chemical Engineering
    Physical Chemistry
  • 出版者:Springer Netherlands
  • ISSN:1572-9567
文摘
The growing demand of industry for traceable temperature measurements up to \(3000\,^{\circ }\hbox {C}\) encourages improvement of calibration techniques for industrial-type radiation thermometers in this temperature range. High-temperature fixed points can be used at such high temperatures, but due to the small diameter of apertures of their cavities (\({\sim }\)3 mm), they are not adapted for the large field-of-views commonly featured by this kind of radiation thermometers. At LNE-Cnam, a Thermo Gauge furnace of 25.4 mm source aperture diameter is used as a comparison source to calibrate customers-instruments against a reference radiation thermometer calibrated according to the ITS-90 with the lowest uncertainties achievable in the Laboratory. But the furnace blackbody radiator exhibits a large temperature gradient that degrades its effective emissivity, and increases the calibration uncertainty due to the lack of information on the working spectral band of the industrial radiation thermometer. In order to estimate the corrections to apply, the temperature distribution (radial and on-axis) of the Thermo Gauge furnace blackbody radiator was characterized and the effective emissivity of the Thermo Gauge cavity was determined by three different methods. Because of this investigation, the corrections due to different fields of view and due to the different spectral bands of the reference pyrometer and the customer’s pyrometer were obtained and the uncertainties on these corrections were evaluated. Keywords Blackbody source Emissivity High temperature Radiation thermometer calibration technique Thermal gradient