Development of multi-planar dose verification by use of a flat panel EPID for intensity-modulated radiation therapy
详细信息    查看全文
  • 作者:Yuji Nakaguchi (1) (3)
    Fujio Araki (2)
    Tomohiro Kouno (3)
    Takeshi Ono (2)
    Kazunari Hioki (3)
  • 关键词:EPID ; Quality assurance ; Verification ; Three ; dimensional measurement ; IMRT
  • 刊名:Radiological Physics and Technology
  • 出版年:2013
  • 出版时间:January 2013
  • 年:2013
  • 卷:6
  • 期:1
  • 页码:226-232
  • 全文大小:975KB
  • 参考文献:1. L茅tourneau D, Gulam M, Yan D, Oldham M, Wong JW. Evaluation of a 2D diode array for IMRT quality assurance. Radiother Oncol. 2004;70:199鈥?06. CrossRef
    2. Greer PB, Vial P, Oliver L, Baldock C. Experimental investigation of the response of an amorphous silicon EPID to intensity modulated radiotherapy beams. Med Phys. 2007;34:4389鈥?8. CrossRef
    3. Arjomandy B, Sahoo N, Ding X, Gillin M. Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance. Med Phys. 2008;35:3889鈥?4. CrossRef
    4. Masi L, Casamassima F, Doro R, Francescon P. Quality assurance of volumetric modulated arc therapy: evaluation and comparison of different dosimetric systems. Med Phys. 2011;38:612鈥?1. CrossRef
    5. Vieira SC, Dirkx ML, Pasma KL, Heijmen BJ. Dosimetric verification of X-ray fields with steep dose gradients using an electronic portal imaging device. Phys Med Biol. 2003;48:157鈥?6. CrossRef
    6. Warkentin B, Steciw S, Rathee S, Fallone BG. Dosimetric IMRT verification with a flat-panel EPID. Med Phys. 2003;30:3143鈥?5. CrossRef
    7. Zeidan OA, Li JG, Ranade M, Stell AM, Dempsey JF. Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging device. Med Phys. 2004;31:463鈥?6. CrossRef
    8. McNutt TR, Mackie TR, Reckwerdt P, Paliwal BR. Modeling dose distributions from portal dose images using the convolution/superposition method. Med Phys. 1996;23:1381鈥?2. CrossRef
    9. Van Elmpt WJ, Nijsten SM, Dekker AL, Mijnheer BJ, Lambin P. Treatment verification in the presence of inhomogeneities using EPID-based three-dimensional dose reconstruction. Med Phys. 2007;34:2816鈥?6. CrossRef
    10. Steciw S, Warkentin B, Rathee S, Fallone BG. Three-dimensional IMRT verification with a flat-panel EPID. Med Phys. 2005;32:600鈥?2. CrossRef
    11. Berry SL, Sheu RD, Polvorosa CS, Wuu CS. Implementation of EPID transit dosimetry based on a through-air dosimetry algorithm. Med Phys. 2012;39:87. CrossRef
    12. Nakaguchi Y, Araki F, Maruyama M. Comparison in dose distributions between 2D-arrarys for pre-treatment IMRT QA. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2012;68(4):443鈥?2. CrossRef
    13. Kawrakow I, Rogers DWO. The EGSnrc code system: Monte Carlo simulation of electron and photon transport. National Research Council of Canada Report PIRS-701. 2009.
    14. Rogers DWO, Faddegon BA, Ding GX, Ma CM, We J, Mackie TR. BEAM: a Monte Carlo code to simulate radiotherapy treatment units. Med Phys. 1995;22:503鈥?4. CrossRef
    15. Roger DWO, Walters B, Kawrakow I. 鈥淏EAMnrc user鈥檚 manual,鈥?National Research Council of Canada Report PIRS-0509 (a) Rev K, 2009.
    16. Walters B, Kawrakow I, Rogers DW. DOSXYZnrc user鈥檚 manual. National Research Council of Canada Report PIRS-794revB 2005.
    17. Zhu XR, Yoo S, Jursinic PA, Grimm DF, Lopez F, Rownd JJ, Gillin MT. Characteristics of sensitometric curves of radiographic films. Med Phys. 2003;30:912鈥?. CrossRef
  • 作者单位:Yuji Nakaguchi (1) (3)
    Fujio Araki (2)
    Tomohiro Kouno (3)
    Takeshi Ono (2)
    Kazunari Hioki (3)

    1. Department of Radiological Technology, Kumamoto University Hospital, 1-1-1 Honjyo, Kumamoto, Japan
    3. Graduate School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Kumamoto, Japan
    2. Faculty of Life Sciences, Kumamoto University, 4-24-1 Kuhonji, Kumamoto, Japan
文摘
Our purpose in this study was to evaluate the accuracy of a new multi-planar dose measurement method. The multi-planar dose distributions were reconstructed at each depth by convolution of EPID fluence and dose kernels with the use of EPIDose software (SunNuclear). The EPIDose was compared with EPID, MapCHECK (SunNuclear), EDR2 (Kodak), and Monte Carlo-calculated dose profiles. The EPIDose profiles were almost in agreement with Monte Carlo-calculated dose profiles and MapCHECK for test plans. The dose profiles were in good agreement with EDR2 at the penumbra region. For dose distributions, EPIDose, EDR2, and MapCHECK agreed with that of the treatment-planning system at each depth in the gamma analysis. In comparisons of clinical IMRT plans, EPIDose had almost the same accuracy as MapCHECK and EDR2. Because EPIDose has a wide dynamic range and high resolution, it is a useful tool for the complicated IMRT verification. Furthermore, EPIDose can also evaluate the absolute dose.