Voltammetric determination of levofloxacin using a glassy carbon electrode modified with poly(o-aminophenol) and graphene quantum dots
详细信息    查看全文
文摘
A strategy was developed for the voltammetric determination of the antibiotic drug levofloxacin (LV) based on a glassy carbon electrode modified with a composite consisting of poly(o-aminophenol) and graphene quantum dots (PoAP/GQD) that was fabricated by electropolymerization. The PoAP/GQD composite provides a large surface area and sensing interface and strongly promotes the oxidation current of LV. Under optimal conditions, the modified GCE displays an oxidation peak current (best measured at a working voltage of 1.05 V vs. SCE) that is linearly related to the levofloxacin concentration in the range from 0.05 to 100 μM, and the detection limit is 10 nM (at an S/N of 3). The method was applied to the determination of levofloxacin in spiked milk samples where is gave recoveries between 96.0 and 101.0 %.
NGLC 2004-2010.National Geological Library of China All Rights Reserved.
Add:29 Xueyuan Rd,Haidian District,Beijing,PRC. Mail Add: 8324 mailbox 100083