智能最佳管电压联合自动管电流在降低胸部CT扫描剂量中的应用
详细信息    查看全文 | 推荐本文 |
  • 英文篇名:Application of the Technique of CARE kV Combined with CARE dose 4D in Reducing Radiation Dosage Ofchest CT Scanning
  • 作者:王俊琴 ; 刘太峰 ; 张海燕
  • 英文作者:WANG Junqin;LIU Taifeng;ZHANG Haiyan;Department of Radiology,Armedpolice Zhejiang Jiaxing Hospital;
  • 关键词:胸部 ; 体层摄影术 ; X线计算机 ; 智能最佳管电压技术 ; 自动管电流 ; 辐射剂量
  • 英文关键词:Chest;;Tomography,X-ray computed;;CARE kV;;CARE Dose 4D;;Radiation dosage
  • 中文刊名:LCFS
  • 英文刊名:Journal of Clinical Radiology
  • 机构:武警浙江省总队嘉兴医院放射科;
  • 出版日期:2019-05-20
  • 出版单位:临床放射学杂志
  • 年:2019
  • 期:v.38;No.346
  • 语种:中文;
  • 页:LCFS201905045
  • 页数:5
  • CN:05
  • ISSN:42-1187/R
  • 分类号:173-177
摘要
目的探讨螺旋CT智能最佳管电压(CARE kV)联合自动管电流(CARE Dose 4D)技术在胸部CT扫描中降低辐射剂量的应用。方法将行胸部CT检查的120例患者分为两组,对照组(n=60)采用CARE Dose 4D扫描技术,试验组(n=60)采用CARE Dose 4D联合CARE kV扫描技术,其余扫描参数和图像重组参数均相同。分析每组患者的肺尖部、主动脉弓下缘水平、左房中部水平及肺底部4幅图像,对比两组的图像质量。测量肺重组及纵隔重组图像的噪声、对比噪声比(CNR)、信噪比(SNR)。记录每组患者的CT容积剂量指数(CTDIvol)、剂量长度乘积(DLP)、有效辐射剂量(ED)。结果试验组ED为(2.82±0.42)mSv,较对照组的(3.19±0.48)mSv明显降低,差异具有统计学意义(P<0.05)。对照组图像平均优良率为93.3%,试验组为91.7%,两组图像优良率相近,差异无统计学意义(P>0.05);然而肺尖水平试验组优良率明显高于对照组,而左房中部水平却低于对照组,主动脉弓下缘及肺底水平两组相近。两组平均CNR、SNR差异均无统计学意义(P>0.05)。结论 CARE kV联合CARE Dose 4D技术,能在有效减低患者辐射剂量的同时,保证图像质量,适合临床推广应用。
        Objective To investigate the combination technique of CARE kV with CARE dose 4 D in reducing radiation dosage of chest CT scanning. Methods 120 patients were randomly divided into two groups who will take chest CT examination,control group(n=60) with CARE Dosed 4 D scanning and research group(n=60) with by CARE dosed 4 D and CARE kV scanning,the rest of the scanning parameters and image reconstruction were set in the same way.Four levels of each CT image from 120 patients were selected,including apex of lung level,central left aorta level,edge of the aortic arch and lung base levels,to compare the image quality. Moreover,contrast to noise ratio(CNR) and signal-to-noise ratio(SNR) were also measuredin lung and mediastinum reconstruction image.Volume CT dose index(CTDI vol) and dosed length product(DLP) of each patient were recorded and effective dosed(ED) was then calculated to evaluate radiation dosage of each group. Results The average ED of the research group was(2.82±0.42) mSv,which was lower than control group(3.19±0.48)mSv with significant difference between two group(P<0.05). Though there was no statistical difference in excellent rate between two group(93.3%vs 91.7%,P>0.05),CT image of research group of lung apex level showed higher excellent rate than that in control group(P<0.05),but CT image in central left aorta level presented lower excellent rate than that of control group(P<0.05),excellent rates of CT images in edge of the aortic arch and base lung levels were similar in the two groups(P>0.05). In addition,the average of CNR,SNR,CTDI vol,DLP and lesion detection rate of the two groups displayed no significant statistical differences(P>0.05). Conclusion The use of technology of CARE kV combined with CARE dose 4 D can not only effectively reduce the radiation dose of patients,but keep the image quality,which is suitable for clinical application.
引文
1刘士远,于红.积极推进胸部低剂量CT扫描的临床应用[J].中华放射学杂志,2010,44:6-7.
    2 Slovis TL.CT and computed radiography:the pictures are great,but is the radiation dose greater than required?[J].AJR,2002,179:39-41.
    3 Hein E,Rogalla P,Klingebiel R,et al.Low-dose CT of the paranasal sinuses with eye lens protection:effect on image quality and radiation dose[J].Eur Radiol,2002,12:1693-1696.
    4李升,邵国良.多层螺旋CT胸部低剂量扫描对肺结节的鉴别诊断分析[J].中国全科医学,2011,14:2823-2825.
    5陆锐,何健,周科峰,等.自动管电流调制扫描及ASIR重建下胸部CT噪声指数的优化[J].医学影像学杂志,2016,26:1012-1016.
    6胡敏霞,赵心明,宋俊峰,等.64层螺旋CT腹部扫描参数优化的个体化选择[J].中华放射学杂志,2012,46:624-628.
    7吴爱琴,郑文龙,许崇永,等.CARE Dose 4D技术降低成人胸部扫描剂量的临床价值[J].中华放射医学与防护杂志,2011,31:98-101.
    8 Winklehner A,Goetti R,Baumueller S,et al.Automated attenuationbased tube potential selection for thoracoabdominal computed tomography angiography:improved dose effectiveness[J].Invest Radiol,2011,46:767-773.
    9 Eller A,May MS,Scharf M,et al.Attenuation-based automatic kilovolt selection in abdominal computed tomography:effects on radiation exposure and image quality[J].Invest Radiol,2012,47:559-565.
NGLC 2004-2010.National Geological Library of China All Rights Reserved.
Add:29 Xueyuan Rd,Haidian District,Beijing,PRC. Mail Add: 8324 mailbox 100083
For exchange or info please contact us via email.